Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats | Text Page (2024)

Summary

Here, we describe a setup for simultaneous recording of electrocardiography and intra-arterial blood pressure (BP) in experimental rats, which can be done with standard equipment in animal facilities and can be applied to physiological or pharmacological studies to investigate pathogenic or therapeutic mechanisms in cardiovascular medicine.

Abstract

For studies related to cardiovascular physiology or pathophysiology, blood pressure (BP) and electrocardiography are basic observational parameters. Research focusing on cardiovascular disease models, potential cardiovascular therapeutic targets or pharmaceutical agents requires assessment of systemic arterial pressure and heart rhythm changes. In situations where radio telemetry systems are not available or affordable, the technique of femoral artery cannulation is an alternative way to obtain intra-arterial pressure waveform recordings and systemic BP measurements. This technique is economical and can be performed with standard equipment in animal facilities. However, invasive arterial pressure recording requires cannulation of small arteries, which can be a challenging surgical skill. Here, we present step-by-step protocols for femoral artery cannulation procedures. Key procedures include the calibration of the data acquisition system, tissue dissection and femoral artery cannulation, and setup of the arterial cannulation system for pressure recording. Surface electrocardiography recording procedures are also included. We also present examples of BP recordings from normotensive and hypertensive rats. This protocol allows reliable direct recordings of systemic BP with simultaneous electrocardiography.

Introduction

Blood pressure (BP) and electrocardiography (ECG) are basic parameters for cardiovascular physiology and medicine. Experimental animal models have been widely applied in biomedical research for various cardiovascular diseases such as hypertensive heart failure1 and procedures for ECG recording and BP measurement can be performed in experimental rats.

There are three methods for BP measurement in rats: intra-arterial cannulation (invasive)2, tail cuff plethysmography (noninvasive)3, and radio telemetry (invasive). The reliability of BP measurement by tail cuff plethysmography can be affected by animal handling during the recording. For example, the tail cuff underestimates the core BP changes that occur simultaneously during the restraint and measurement phases4. Radio telemetry is considered the best "gold standard" technique for monitoring BP and heart rate in awake and freely moving animals5. However, since radio telemetry hardware and software are costly, intra-arterial cannulation is also widely used as an economical alternative.

Intra-arterial cannulation requires considerable microsurgical skill but yields the real waveforms of arterial pressure. BP can be recorded through a saline-filled catheter inserted in the radial, femoral, or brachial artery. This method of direct invasive BP measurement requires pre-surgical animal preparation, anesthesia, immobilization of laboratory animals, surgical skill in tissue dissection and arterial cannulation, and proper calibration before acquiring the measurement.

Rodent surface ECG is similar to human ECG. A rat ECG has sequences of P waves, QRS complexes, T waves, and QT intervals6. The P wave, PR interval, QRS complex, and T waves reflect atrial depolarization, impulse conduction from the atrial to the AV node, ventricular depolarization, and repolarization, respectively. The QT interval is defined as the period from the initiation of the Q wave to the end-point of the T wave where it returns to the iso-electrical baseline1.

The ECG indicates the cardiac systole and diastole phases; therefore, the simultaneous recording of the surface ECG correlates with the invasive BP measurement. By using a combination of methodologies, it is possible to elucidate pathophysiological changes in a disease model or the pharmacological effects of a drug or therapy in cardiovascular medicine.

A spontaneous hypertensive rat (SHR) strain had been obtained by inbreeding of Wistar rats with high BP in Japan. The BP rises from 5 to 10 weeks of age and becomes stationary from 30 to 35 weeks of age7. Wistar-Kyoto rats (WKY) have systolic BP about 130 mmHg7 and are commonly used as normo-tensive control. We used SHR and WKY to demonstrate the result of intraarterial cannulation BP and ECG recording.

Subscription Required. Please recommend JoVE to your librarian.

Protocol

All the animal experiments described were approved by the Institutional Animal Care and Use Committee of Kaohsiung Medical University.

1. Animal Care

  1. To avoid difficult cannulation due to small arterial size, use rats with body weight over 200 g.
  2. Remove chow from the cage and fast rats overnight.
  3. Offer water ad libidum unless there is a special experimental design.

2. Experimental Preparation

  1. Obtain the following materials: forceps with teeth (Figure 1A), surgical scissors (Figure 1B), forceps with fine tips (Figure 1C, 1D), forceps with angled tips (Figure 1E), bulldog vascular clamp (Figure 1F), silk string approximately 20 cm in length (Figure 1G), micro scissors (Figure 1H), intra-arterial cannula, a sterile polyethylene (PE) tube with an internal diameter of 0.5 mm and an outer diameter of 0.9 mm, 25–30 cm in length, connected to a 26 G x 1/2’’ needle (Figure 1I), two three-way stopco*cks (Figure 1J) for connecting the intra-arterial cannula and the pressure transducer (Figure 1K), and 1 mL syringes filled with heparinized saline (100 IU/mL).
  2. Animal Preparation
    1. Anesthetize the rat by inhalation of isoflurane (placing the rats in a 25 cm x 25 cm x 14 cm induction chamber saturated with 4% isoflurane, followed by nosecone use with 3% isoflurane).
      NOTE: If the animal is not sufficiently anesthetized, turn up the flow rate of isoflurane.
    2. Test the pain reflex by pinching the toes.
    3. Place the rat in a supine position on a pre-cut Styrofoam board (or thick cardboard). Fix the four legs with rubber bands to immobilize the body (Figure 2A).
      NOTE: To avoid potential noise during the ECG recording, the placement surface must not be electrically conductive.
  3. Prepare instruments for BP and ECG recording. Include an analog input unit to acquire signals, a pressure transducer with a compatible hub, three bipolar needle-tipped electrocardiogram leads, and a computer with suitable software.

3. Pressure Transducer Calibration

  1. Before initiation of BP recording, calibrate with a standard mercury sphygmomanometer (Figure 1L).
  2. Remove the pressure cuff from the sphygmomanometer and connect the three-way stopco*ck on the inflation tube to the pressure transducer (Figure 1K) of the data acquisition system.
  3. Screw the air-release valve clockwise. Keep eyes on the gauge and keep pumping the inflation bulb. When the gauge shows 100 mmHg, switch the three-way stopco*ck to connect to the pressure transducer. Use 100 mmHg pressure for the calibration. The conversion factor for calculating BP will be determined automatically.
  4. Release the pressure by screwing the air-release valve counterclockwise until the pressure of the sphygmomanometer is back to zero.
  5. Repeat step 3.2 with pressure of 200 mmHg.
  6. Detach the pressure transducer from the mercury sphygmomanometer.
  7. Connect the pressure transducer to the three-way stopco*ck of the PE catheter (Figure 1I).

4. Mini-surgery for Cannulation of the Femoral Artery

  1. Surface landmark identification and excision of the skin (Figure 2)
    1. Identify the location of inguinal crease (indentation at the junction between abdomen and thigh) (dashed line in Figure 2A).
    2. Pinch the full layer of skin at the center of the inguinal crease.The fur may be shaved or removed from the incision site using electric clippers or a depilatory cream prior to incision.
    3. Lift the skin and cut it off with the surgical scissors at an orientation approximately parallel to the ipsilateral thigh (Figure 2B). The femoral nerve and vessels are underneath the exposed subcutaneous tissue (Figure 2C).
  2. Dissection of the tissue to expose the femoral artery
    1. Dissect the tissue using forceps with fine tips, layer by layer. Stop the dissection at the level of femoral vessels. Make sure the dissection is not injuring the vessels underneath.
    2. Carefully use forceps (Figure 1C or 1D) to clear out the soft tissue along the femoral nerve and vessels to obtain good observation. The nerve is fiber-like in texture. The vein is dark purple and the artery is pulsatile (Figure 3A).
    3. Use forceps with angled tips (Figure 1E) to extend the exposed length of the femoral artery and vein (Figure 3B).
  3. Cannulation of the femoral artery (Figure 3)
    1. Use forceps to separate the femoral vein from the artery and apply a bulldog clamp at the femoral artery as cranially as possible (Figure 3C).
    2. Make a loose tie of the two silk strings, one just below the bulldog clamp and the other at the caudal terminal of the exposed femoral artery (Figure 3D).
    3. Make a small hole over the ventral side of the femoral artery (Figure 3E) using the micro scissors (Figure 1H).
    4. Insert the tip of the PE catheter through the small hole and advance the catheter cranially.
      NOTE: Do not apply torsion while advancing the PE catheter. The torsion force may twist the femoral artery and result in luminal stenosis.
    5. Remove the bulldog clamp after the PE catheter is securely advanced into the femoral artery lumen.
      NOTE:Keep eyes on the PE catheter while removing the bulldog clamp from the femoral artery. The observation of a backflush of blood into the PE catheter associated with observation for pulse confirms its placement in the arterial lumen.
    6. Tighten the upper silk string to secure the position of the PE catheter.
      NOTE: Any accidental pull or dislocation of the PE catheter can result in major bleeding.
    7. Tighten the lower silk string to prevent bleeding from the caudal side of the femoral artery.
  4. Confirmation of the success of femoral artery cannulation
    1. Use a 1 mL syringe to inject 0.1–0.2 mL of heparinized saline into the femoral artery. The resistance on injection must be trivial. If any obvious resistance upon injection is noticed, check the whole cannulation again.
      NOTE: Make sure the three-way stopco*cks are appropriately switched before applying any negative pressure or saline injection into the femoral artery cannula. At this time point, any injection into the pressure transducer will need a re-calibration.
    2. Check if there is any oozing around the cannulation site. If not, cover the surgical site with a wet cotton ball.

5. Recording of Blood Pressure

  1. After a smooth flush for the PE catheter, attach the PE catheter three-way stopco*ck to the one on the pressure transducer (Figure 4).
  2. Make sure that there are no air bubbles in the cannulation system. Also check the connection junctions of the three-way stopco*ck.
  3. Start the data acquisition system with sampling frequency of 1,000 Hz to record the BP. Arterial pressure waves will be demonstrated (Figure 6).
  4. Allow the whole setup to stabilize for at least 3-5 min. In cases with unsteady signals, stabilization time can be extended to 15 min.
  5. Check the cannulation site periodically to make sure there is no bleeding.
    NOTE: When acquiring data from a pressure transducer, it is important to place the transducer at the level of the animal’s heart.

6. Surface ECG

  1. Check the three leads of the bipolar ECG to make sure the positive, negative, and reference platinum electrodes are intact.
  2. Insert the leads subcutaneously at the left foreleg, right foreleg, and right hindleg (Figure 5).
  3. Attach the electrode hubs to a custom-built ECG amplifier with sampling frequency of 1,000 Hz and filter frequency of 3–500 Hz. Keep ECG leads steady during recording. Motion of the ECG leads can produce unsteady baselines and artifacts.

7. Animal Euthanasia After Completing of the Experiment

  1. After completing of the BP and ECG recording, stop the acquisition system. Remove the electrodes. Ligate the femoral artery by tightening up the previously placed silk string right after withdrawing the PE catheter.
  2. Place the rats individually in the visible euthanasia chamber connected to compressed carbon dioxide (CO2) gas cylinders. Seal the top securely.
  3. Introduce 100% CO2 with a fill rate of about 10% to 30% of the chamber volume per minute.
  4. Keep your eyes on the rat; lack of respiration and faded eye color should appear within 2−3 min.
    NOTE: If cessation of respiratory movementdoes not occur after 3 min, the system should be examined for the chamber fill rate, CO2 supply, or leaks.
  5. Maintain CO2 flow to the chamber for 1 more minute after respiration ceases.
  6. Ascertain cardiac and respiratory arrest and note fixed and dilated pupils to confirm death.
    NOTE: If the rat is not dead but in CO2 narcosis, use a secondary method of euthanasia such as bilateral thoracotomy.

Subscription Required. Please recommend JoVE to your librarian.

Representative Results

We purchased SHR and normotensive Wistar-Kyoto WKY rats from the National Laboratory Animal Center (Taipei, Taiwan). All animals were housed in a temperature-controlled facility (20−22 °C) with free access to water and standard chow on a 12 h light/dark cycle.

We used six 47-week-old rats and they were weighed before the BP and ECG measurement. The representative tracings from simultaneous recording of ECG and BP in SHR and WKY are shown in Figure 6. Table 1 shows parameters for BP and heart rate. Statistical (t) tests revealed significantly higher systolic BP and mean BP in SHR (124.5 mmHg ± 15.1 mmHg and 84.3 mmHg ± 5.0 mmHg) than in WKY (90.0 mmHg ± 7.5 mmHg and 67.5 mmHg ± 5.0 mmHg) (P < 0.05) (Table 1).

The parameters of P wave, PR intervals, QRS width, and QT intervals can be measured from the ECG recordings (Table 2). Choice of right vs. left femoral cannulation in relation to ECG electrode placement (at the right hind limb) did not affect the BP or ECG signals.

Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats | Text Page (1)
Figure 1: Materials. (A) Forceps with teeth, (B) surgical scissors, (C) tissue forceps, (D) tissue forceps with fine tips, (E) angle tip forceps, (F) bulldog vascular clamp, (G) silk string, (H) micro scissors, (I) polyethylene catheter connecting with a 26 G x 1/2'' needle and a three-way stopco*ck (J), (K) pressure transducer connecting with a three-way stopco*ck, (L) mercury sphygmomanometer with a stopco*ck (green arrow) and inflation bulb with an air-leak valve (white arrow). Please click here to view a larger version of this figure.

Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats | Text Page (2)
Figure 2: Surface landmark for femoral artery dissection. (A) Supine rat with left inguinal crease highlighted with a dashed line. (B) Groin skin picked up by forceps with teeth to be cut off by surgical scissors. (C) The surgical zone for femoral artery dissection. Please click here to view a larger version of this figure.

Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats | Text Page (3)
Figure 3: Tissue dissection and cannulation of the femoral artery. (A) Exposed femoral nerve (yellow arrow), femoral vein (blue arrow), and femoral artery (red arrow) after tissue dissection. (B) Femoral vein and artery after the nerve cut-off. (C) Application of a bulldog vascular clamp over the cranial terminal of the dissected femoral artery. (D) Femoral artery with a bulldog clamp and two silk strings with loose ties. (E) A small hole on the ventral side of femoral artery, created using micro scissors. (F) Insertion of the polyethylene catheter into the femoral artery. Please click here to view a larger version of this figure.

Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats | Text Page (4)
Figure 4: Femoral artery cannulation for pressure recording. The intra-arterial cannula (highlighted by a dashed green line) is connected with two three-way stopco*cks (blue arrows) with syringes filled with heparinized saline. The pressure transducer (red arrow) is linked to the cannulation through the three-way stopco*cks. Please click here to view a larger version of this figure.

Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats | Text Page (5)
Figure 5: The whole experimental setup for invasive femoral arterial pressure and electrocardiography. Three electrocardiographic leads (red arrows) with platinum needle electrodes subcutaneously inserted over bilateral fore legs and right leg (yellow arrows). The exposed right femoral artery with cannulation (blue arrow). Please click here to view a larger version of this figure.

Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats | Text Page (6)
Figure 6: Representative tracings. Simultaneous recordings of ECG and BP in WKY (left) and SHR (right) for the ECG (top), and arterial pressure waves (bottom). Please click here to view a larger version of this figure.

WKYSHRP value
(n = 3)(n = 3)
Weight, g407.7 ± 6.4353.7 ± 10.3<0.01
Heart rate, /min241.3 ± 15.0262.0 ± 15.70.18
Systolic BP, mmHg90.0 ± 7.5124.5 ± 15.1<0.05
Diastolic BP, mmHg56.3 ± 4.064.2 ± 0.60.08
Mean BP, mmHg67.5 ± 5.084.3 ± 5.0<0.05
Pulse pressure, mmHg33.6 ± 4.460.4 ± 15.10.08

Table 1: Blood pressure levels measured by invasive femoral artery cannulation. Data are presented as mean ± standard deviation; SHR = spontaneously hypertensive rats; WKY = Wistar-Kyoto rats; BP = blood pressure.

WKYSHRP value
(n = 3)(n = 3)
Heart rate, /min241.3 ± 15.0262.0 ± 15.70.18
P wave, msec38.7 ± 9.343.0 ± 4.60.52
PR interval, msec59.0 ± 7.861.3 ± 4.70.69
QRS width, msec43.0 ± 11.350.0 ± 2.60.4
QT interval, msec 112.0 ± 12.8116.7 ± 9.50.15

Table 2:Measurements of electrocardiographic parameters. Data are presented as mean ± standard deviation; SHR = spontaneously hypertensive rats; WKY = Wistar-Kyoto rats.

Subscription Required. Please recommend JoVE to your librarian.

Discussion

Invasive arterial cannulation allows highly accurate measurement of BP. It can be done with a PE tube without requiring an expensive catheter. Invasive BP measurement can also be performed simultaneously with a recording of the surface ECG.

The major learning curve for this method is the experimental skill required to cannulate small blood vessels. In experienced hands, the successful rate for femoral artery cannulation can approach 100%. Practice is recommended before performing realexperiments. Some noteworthy points during the procedure: (1) make sure animals are completely anesthetized before starting the surgery; (2) use a bulldog clamp at the high position of the exposed artery before cutting the arterial wall to prevent extensive bleeding; (3) make the incision only over the ventral wall of the artery and avoid damaging the dorsal wall; and (4) steady placement of the cannulation system and electrocardiographic leads helps avoid motion artifacts during the recording.

The potential complications with this procedure include hemorrhage from the surgical dissection. The hemorrhage in a successful and smooth cannulation of an artery is very trivial. Major hemorrhage can be caused by accidental damage to the femoral artery or vein or the major branches during the tissue dissection. The hemorrhage can be stopped by focal compression with a clear cotton ball. Another cause of major hemorrhage is dislocation of the femoral cannulation. In this case, the blood loss is usually large enough to result in a significant drop in systemic BP. At this point, BP recording can be terminated early.

Telemeter implantation also is an established technique for invasive BP recording8. The advantage of telemetry is the ability to obtain a high-fidelity continuous recording over relatively long periods of time in conscious, freely moving animals without the limitations of restraint or anesthesia. Direct BP recordings can be done with successful subcutaneous implantation of radio-transmitters and carotid artery cannulation9. However, it can be challenging to implant the telemeter successfully and the hardware and software for radio telemetry are costly.

There are some limitations for invasive BP measurement by femoral artery cannulation. First, it cannot be performed in unrestrained and conscious rats. Second, BP in the femoral artery may be higher or lower than BP in the central aorta. Third, arterial cannulation can be difficult for small rodents, where femoral arteries may be too small to insert the PE tube. Fourth, the cannulated artery should be ligated after completing the BP recording and ligation of femoral artery will result in hind leg ischemia. Due to this limitation, the animals are usually euthanized following invasive BP recording.

Invasive femoral artery cannulation requires restraining and anesthetizing the animals, which potentially introduces stress and influences both BP and electrocardiography data10. The stress can be alleviated by appropriate anesthesia. Anesthesia includes injection or inhalation protocols. In our past experience, high elevated BP in I.P. anesthetized WKY rats suggests that the animals were still under stress or in pain. Normal BP in conscious WKY rats is 130/80 mmHg; this dropped to below 100 mmHg in our inhalation-anesthetized WKY rats (Table 1). Inhalational anesthesia (isoflurane) has several advantages over injectable agents: quick onset, minimal animal handling, ease of control, no controlled drugs, and a quick recovery. The disadvantages are the cost of the equipment, the hazard of human exposure in case with substantialgas leakinto the working environment, and a significant suppressive effect on BP. Isoflurane-induced BP reduction should be considered when choosing the anesthesia regime.

Further applications are possible for invasive BP measurement. According to the fluid dynamics principles for pulse wave velocity (PWV), the stiffer artery propagates the pulse wave faster11. When applying double cannulations over the carotid artery and femoral artery, the aortic PWV can be determined. The length of aortic pulse propagation can be measured as the distance between the tips of the cannula at the carotid (proximal) and distal (femoral) arteries. The PWV is the ratio between the aortic length and the difference between the time at the minimal values of the proximal and distal arterial pulses12.

The setup described above for simultaneous electrocardiography and intra-arterial BP recording in experimental rats is an inexpensive and highly accessible technique for physiological or pharmacological studies investigating pathogenic or therapeutic mechanisms in cardiovascular medicine.

Subscription Required. Please recommend JoVE to your librarian.

Disclosures

The authors have nothing to disclose.

Acknowledgments

This study was supported by Taiwan Ministry of Science and Technology grants MOST 104-2314-B-037-080-MY3 and MOST 107-2314-B-037-110 to HCL and Taiwan National Health Research Institutes grant NHRI-EX107-10724SC.

Materials

NameCompanyCatalog NumberComments
Polyethylene tubeBECTON DICKINSON427401internal diameter of 0.5 mm, outer diameter of 0.9 mm
26 G x 1/2"" needleTERUMO160426D
Adson ForcepsTOP Line12-54012 cm (4.75") Long, Straight, 1 x 2 Teeth
Bulldog vascular clampTeleflex3575818 mm
ComputerAUSUSX453M
Exernal analog signal recording deviceiWorxT5141538This allows the recording of up to three channels of ECG, EMG or EEG as well as GSR (skin conductance) from a single iWire input on the recording Module.
Graefe ForcepsAESCULAP Surgical InstrumentsBD312RMICRO DRESSING FORCEPS, CURVED, SERRATED, 105 mm, 4 1/8
Mecury sphygmomanometerSpiritCK-101
Pressure transduceriWorxIworxBP100
Semken ForcepsMEDE TECHNIK10-104100 mm
SoftwareLabScribe3
Surgical scissorsHEBU171414.5 cm long
Syringe (1 mL)TERUMO160426D
Three-way stopco*cksCole-ParmerEW-30600-23
Tipped forcepsWorld Precision Instruments50450611 cm long, 0.1 x 0.06 mm Tips
Vannas ScissorsWorld Precision Instruments5000868.5 cm long, Straight, 0.025 x 0.015 mm Tips, 7mm super fine Blades

DOWNLOAD MATERIALS LIST

References

  1. Shiou, Y. -L., Huang, I. C., Lin, H. -T., Lee, H. -C. High fat diet aggravates atrial and ventricular remodeling of hypertensive heart disease in aging rats. Journal of the Formosan Medical Association. , (2017).
  2. Parasuraman, S., Raveendran, R. Measurement of invasive blood pressure in rats. Journal of Pharmacology, Pharmacotherapeutics. 3 (2), 172-177 (2012).
  3. Fink, G. D. Does Tail-Cuff Plethysmography Provide a Reliable Estimate of Central Blood Pressure in Mice? Journal of the American Heart Association: Cardiovascular and Cerebrovascular Disease. 6 (6), e006554 (2017).
  4. Wilde, E., et al. Tail-Cuff Technique and Its Influence on Central Blood Pressure in the Mouse. Journal of the American Heart Association: Cardiovascular and Cerebrovascular Disease. 6 (6), (2017).
  5. Braga, V. A., Prabhakar, N. R. Refinement of Telemetry for Measuring Blood Pressure in Conscious Rats. Journal of the American Association for Laboratory Animal Science : JAALAS. 48 (3), 268-271 (2009).
  6. Kaese, S., Verheule, S. Cardiac electrophysiology in mice: a matter of size. Frontiers in Physiology. 3 (345), (2012).
  7. Okamoto, K., Aoki, K. Development of a strain of spontaneously hypertensive rats. Japanese Circulation Journal. 27, 282-293 (1963).
  8. Van Vliet, B. N., Chafe, L. L., Antic, V., Schnyder-Candrian, S., Montani, J. P. Direct and indirect methods used to study arterial blood pressure. Journal of Pharmacological and Toxicological Methods. 44 (2), 361-373 (2000).
  9. Alam, M. A., Parks, C., Mancarella, S. Long-term Blood Pressure Measurement in Freely Moving Mice Using Telemetry. Journal of Visualized Experiments. (111), e53991 (2016).
  10. Irvine, R. J., White, J., Chan, R. The influence of restraint on blood pressure in the rat. Journal of Pharmacological and Toxicological Methods. 38 (3), 157-162 (1997).
  11. Nvd Vosse, F., Stergiopulos, N. Pulse Wave Propagation in the Arterial Tree. Annual Review of Fluid Mechanics. 43 (1), 467-499 (2011).
  12. Cosson, E., et al. Aortic stiffness and pulse pressure amplification in Wistar-Kyoto and spontaneously hypertensive rats. The American Journal of Physiology: Heart and Circulatory Physiology. 292 (5), H2506-H2512 (2007).

Tags

Simultaneous Electrocardiography Recording Invasive Blood Pressure Measurement Rats Cardiovascular Disease Models Cardiovascular Therapeutic Targets Pharmaceutical Agents Femoral Artery Cannulation Intraarterial Pressure Waveforms Animal Facility Hsin-Ting Lin Technician Anesthetized Rat Supine Position Surgical Board Lack Of Response To Toe Pinch Rubber Bands Blood Pressure Instruments Electrocardiogram Instruments Analog Input Unit Pressure Transducer Compatible Hub Bipolar Needle-tipped Electrocardiogram Leads Computer With Software Blood Pressure Recording Pressure Cuff Sphygmomanometer Three-way Stopco*ck Data Acquisition System

Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats | Text Page (2024)

FAQs

What is the invasive method of blood pressure measurement in rats? ›

A small incision (1.5–2 cm) is made in the neck of the rat for tracheostomy and carotid artery cannulation. The skin in the neck region is carefully cut open, and a slit incision is made in the rat platysma muscles.

How to measure rat blood pressure? ›

The most common techniques currently employed for monitoring blood pressure (BP) in conscious rats and mice are the tail cuff and the exteriorized catheter that feeds a pressure transducer located outside the cage.

What is the normal blood pressure for a rat? ›

When blood pressure in both groups became relatively stable during the study period, the average systolic, mean and diastolic pressure were 124, 101 and 82 mmHg respectively for the male rats. The average pressures for the female rats were 121 (systolic), 99 (mean) and 80 (diastolic) mmHg (Figures 1 & 2).

What is the tail cuff method? ›

This method uses a specialized volume pressure recording (VPR) sensor, and measures blood volume changes that are placed over the animal's tail. Mice do need to be restrained in specific holders and artificially heated to maintain normal BP.

What is invasive method of blood pressure measurement? ›

Arterial pressure monitoring is a form of invasive blood pressure monitoring and is done through the cannulation of a peripheral artery. This form of monitoring is commonly utilized in the management of critically ill and perioperative patients.

What is the difference between noninvasive blood pressure monitoring and invasive blood pressure monitoring techniques? ›

Since non-invasive monitors measure blood pressure by volume displacement or flow detection and invasive ones measure pressure impulses rather than flow, it was concluded that the pressure measured by the non-invasive monitor more accurately reflects the propulsive pressure-causing flow when inotropic pressure pulse ...

What is the mean arterial pressure in rats? ›

MAP(% CONTROL)

(C) A verage frequency distributions of arterial pressure over a 24-hour period in the control and denervated groups of rats. 100% of control mean arterial pressure in both groups is 119 mm Hg.

What is the normal pulse rate for a rat? ›

Table 1
SpeciesBody Weight (kg)Heart Rate (bpm)
Rat0.225–0.52250–493
Rabbit1–6130–300
Canine7–1670–160
Sheep20–16060–120
3 more rows
Oct 15, 2013

What is nibp for rats? ›

The NIBP System includes a specialized tail cuff and pulse transducer, used for intermittent mouse and rat blood pressure measurement based on the periodic occlusion of tail blood flow.

Can rats have high blood pressure? ›

Studies on the Kyoto (SHR) and the New Zealand (GHR) strains of genetically predisposed hypertensive rats have shown that in the SHR neurogenic influences, primarily of higher central origin, play an important role in the initiation of hypertension.

Which animal has the highest blood pressure in the world? ›

The giraffe has one of the highest blood pressures of any mammal. It reaches pressures of up to 280/180 mm Hg, which is twice that found in humans (120/80 mmHg).

What is normal cardiac output in rats? ›

Cardiac output of our normal rats averaged 48 it 4 ml./min.

How do you position a cuff? ›

It is recommended that the cuff should be wrapped around the upper arm with the midline of the bladder placed over the brachial artery during blood pressure (BP) measurement.

What is the cuff method for blood pressure? ›

Stretch out your arm, palm upward. Place the cuff on your bare upper arm one inch above the bend of your elbow. Make sure the tubing falls over the front center of your arm so that the sensor is correctly placed. Pull the end of the cuff so that it's evenly tight around your arm.

How do you measure blood pressure in mice? ›

The 2 major options available for measuring murine blood pressure are noninvasive volume and pressure tail-cuff systems and implantable telemetry devices.

What is the invasive procedure for high blood pressure? ›

Renal denervation (also called renal ablation) is a minimally invasive, investigational procedure to treat hypertension (high blood pressure) that hasn't improved with other treatments. This type of hypertension is called resistant hypertension.

How is invasive arterial blood pressure measured? ›

The usual method of invasive blood pressure monitoring consists of the percutaneous insertion of a small-bore (18- to 22-gauge) plastic catheter into a peripheral artery. The catheter is physically connected via high-pressure plastic tubing to an electronic pressure transducer and display unit.

What method is used to collect blood from rats? ›

Blood is collected using a 23 gauge needle from the lateral saphenous vein. Blood flow is stopped by applying pressure with sterile gauze to achieve hemostasis. No more than four blood samples should be taken within any 24-hour period. For multiple samples the scab or the blood clot should be removed.

What is a non-invasive blood pressure system for rodents? ›

The NIBP System includes a specialized tail cuff and pulse transducer, used for intermittent mouse and rat blood pressure measurement based on the periodic occlusion of tail blood flow. Tail cuffs are available for either mice or rats. Suitable rodent restrainers are available separately.

References

Top Articles
Latest Posts
Article information

Author: Pres. Lawanda Wiegand

Last Updated:

Views: 5893

Rating: 4 / 5 (51 voted)

Reviews: 90% of readers found this page helpful

Author information

Name: Pres. Lawanda Wiegand

Birthday: 1993-01-10

Address: Suite 391 6963 Ullrich Shore, Bellefort, WI 01350-7893

Phone: +6806610432415

Job: Dynamic Manufacturing Assistant

Hobby: amateur radio, Taekwondo, Wood carving, Parkour, Skateboarding, Running, Rafting

Introduction: My name is Pres. Lawanda Wiegand, I am a inquisitive, helpful, glamorous, cheerful, open, clever, innocent person who loves writing and wants to share my knowledge and understanding with you.